Type
|
Karl Fischer Moisture Titrator
|
Model
|
MKC-710M
|
MKC-710S
|
Product configuration
|
MCU-710M+MKC-710+IDP-100+Automatic Solvent Change unit
|
MCU-710S+MKC-710+IDP-100+Automatic Solvent Change unit
|
Measuring method
|
Karl Fischer Coulometric titration
|
Measuring range
|
Water content / Bromine index : 10ug to 300mg (depends on reagent)
|
Measurement cell
|
2-Component or 1-Component
|
Precision
|
Relative standard deviation : less than 0.3% (n=10)
*Per KEM standard measurement conditions and standard liquids
|
Display resolution
|
0.1ug
|
Control method
|
Constant current pulse time control
|
Endpoint detection
|
Alternate current polarization method with a twin platinum electrode
|
EP sense method
|
Selective drift stability or limit measurement time
|
Required solvent
|
2-Component cell : Anolyte 100mL , Catholyte 5mL
1-Component cell : Anolyte 150mL
|
Method
|
120
|
Key operation
|
Touch panel
|
Displays
|
1) 8.4-inch color LCD 800 x 600 dots
|
2) English / Japanese / Mandarin Chinese / Korean / Russian / Spanish / German / French
|
3) Simultaneous 4-channel display (Can also display Automatic Potentiometric Titrator simultaneously)
|
3) 1-channel display
|
Calculation
|
Concentration of water content , statistics data processing (mean , SD and RSD) and automatic averaging of blank value and factor value
|
Data storage
|
500 samples
|
GLP conformance
|
Registration of operator / User group administration
Check performance with standard substance :
Reminder of scheduled check date / Record of check results
Reagent life control : Reminder of expiration/Reminder reagent replacement date
Management of conduction time : Display of operating time
|
External I/O
|
RS-232C port x4 :
for Dot matrix printer , Electronic balance , Data Capture Software (SOFT-CAP) , Evaporator , Multiple sample changer
USB ×1 :
for USB flash drive , Thermal printer , A4 printer , Keyboard , Barcode reader , Foot switch , USB HUB
|
LAN ×1 : for Personal computer (PC)
|
-
|
Extensibility
|
Measuring instrument :
Automatic Potentiometric Titrator (AT-710) , Karl Fischer Moisture Titrator (MKV-710/MKC-710)
Three of these instruments can be added
|
-
|
Evaporator ADP-611
|
Multiple sample changer : CHK-501
|
Ambient condition
|
1) Temperature
|
|
:
|
|
5 to 35°C
|
2) Humidity
|
|
:
|
|
85%RH or below (no condensation)
|
|
Power source
|
AC100 to 240V +/- 10% 50/60 Hz
|
Power consumption
|
Main unit
|
|
:
|
|
Approx. 30W
|
Printer
|
|
:
|
|
Approx. 7W
|
|
Dimensions
|
Touch-on panel controller
|
|
:
|
|
225(W) x 190(D) x 42(H) mm
|
Titration unit
|
|
:
|
|
141(W) x 292(D) x 244(H) mm
|
Stirrer
|
|
:
|
|
107(W) x 206(D) x 340(H) mm (not incl. Solvent Change unit)
|
Solvent Change Unit
|
|
:
|
|
240(W) x 140(D) x 405(H) mm (not incl. tubing)
|
Printer
|
|
:
|
|
106(W) x 180(D) x 88(H) mm
|
|
Weight
|
Touch-on panel controller
|
|
:
|
|
Approx.1.5kg
|
Titration unit
|
|
:
|
|
Approx.3.0kg
|
Stirrer
|
|
:
|
|
Approx.2.0kg
|
Solvent Change Unit
|
|
:
|
|
Approx.0.6kg
|
Printer
|
|
:
|
|
Approx.0.4kg
|
|
Conformity standard
|
CE marking EMC : EN61326-1 LVD : EN61010-1
|