Density meter and Refractometer Combined Meter DA-6XX+RA-6XX

Density meter and Refractometer Combined Meter DA-6XX+RA-6XX

Model No.︰DA-6XX+RA-6XX

Brand Name︰KEM

Country of Origin︰Japan

Unit Price︰-

Minimum Order︰1 pc

Inquire Now

Product Description

General Concentration Meter <DA-6XX+RA-6XX>

  • The two sensors each for density and refractive index need just small amount of sample liquids in order to simultaneously measure both density (specific gravity) and refractive index.
  • The built-in thermostat controls measurement temperature in a short span of time to preset degree for precise and quick measurements of both density (specific gravity) and refractive index. The measuring units each for density (specific gravity) and refractive index can individually control and preset temperature.
  • Density measuring unit features viscosity correction for samples of higher viscosity, allowing measured density values to be compensated.
  • Measuring range is wide enough to measure any kinds of samples: Density : 0 to 3 g/cm3 Refractive index: 1.32000 to 1.58000 (DA-645+RA-620)
  • The optional USB allows you to expand the versatility of operations including more storage space of measurement parameters and data as well as a user's own concentration conversion tables.
  • The optional data acquisition software designed for DA-6XX+RA-6XX allows you download measurement results to PC so that the data can be further processed according to a user's needs and requirements for easier management of acquired data.
  • The commercially available bar code readers compatible with RS-232C serial interface can be connected to AUX port and conveniently read sample ID and lot number.

Specifications of General Concentration meters

Type and Model name DA-645+RA-600 DA-645+RA-620
Measuring method
Density : Resonant frequency detection
Refractive Index : Detection of critical angle with Na-D line
Measuring range
Density : 0 - 3 g/cm3
R.I. : 1.32 - 1.70
Density : 0 - 3 g/cm3
R.I. : 1.32 - 1.58
Measuring temperature Density: 0 - 96°C, R.I.: 5 - 75°C
Accuracy Density ±0.00005 g/cm3
When calibrated with pure water and dry air
Refractive
index
±0.0001 ±0.00002
Temperature Density: ±0.03°C
Minimum sample required 6ml
Auto viscosity correction Built-in
Measuring time 1 - 4 minutes by manual
Display Density: 5.7-inch, R.I.: 4.7-inch TFT LCD
Sampling Manual with syringe
Automatic by peristaltic pump
Calibration "Air & Water " and "Other STD"
Auto statistical data Mean value; Standard deviation (SD) ; Relative SD
Recalculation Data can be recalculated
Printout IDP-100 impact dot printer (Option)
Data I/O USB flash drive as data soorage medium.
External control
LAN : Personal computer
USB : Keyboard
USB : Barcode reader
RS-232C : Printer
USB: USB flash drive
Power source 100 -  240V AC ; 50/60Hz
Power consumption Approx. 60W

Density / Specific Gravity Meter ︰ Density meter and Refractometer Combined Meter DA-6XX+RA-6XX

Product Image